Eppendorf Multipurpose Centrifuges 5810 & 5810R
POCD Scientific 20% free goods bonus giveaway with 58xx centrifuges
AND a 3 year extended warranty on parts giving you a total of 5 years parts and 2 year labour
|
|
|
|
|
We sell, supply, service and carry spare parts for these products.
All equipment servicing to manufacturer’s specs by Eppendorf factory certified technicians.
Calibration, repair, service, spare parts, Compliance Covering - IQ/OQ/PQ, calibration and validations.
On-site repairs are available for customers in Sydney, Melbourne and Brisbane.
For other areas/States, we offer a workshop service or spare parts.
| overview |
features |
technical |
rotors |
accessories |
downloads |
FAQs |
 POCD Scientific ~ Eppendorf’s Australian distributor for Multipurpose Centrifuges 5810 & 5810R ~ overview |
Eppendorf Multipurpose Centrifuges feature powerful, maintenance-free motors, and they accommodate a variety of rotors for unparalleled application versatility. They are three centrifuges in one: a high capacity general purpose centrifuge for cell harvesting; a high-speed centrifuge for separating cell lysates; and a micro centrifuge for DNA precipitations.
 POCD Scientific ~ Eppendorf’s Australian distributor for Multipurpose Centrifuges 5810 & 5810R ~ features |
- Interchangeable swing-bucket and high-speed fixed-angle rotors spin microcentrifuge tubes, popular glass and plastic tubes, microplates, and blood collection tubes
- Variable speed, maintenance-free motor spins up to 14,000 rpm, 20,800 x g
- Speed can be set from 200 rpm to maximum speed in increments of 10 rpm
- Low profile 11-inch access height for easy sample loading
- Automatic rotor recognition senses rotor type to set maximum allowable speedensures safety
- Simple, easy-to-use programming:
- Time, temperature, speed/g-force, and radius values can be entered quickly
- 34 user-defined programs with write protection for quick parameter entry ensures reproducibility
- Adjustable-speed Short Spin button
- Ten acceleration and braking ratesideal for protecting sensitive samples
Cooling features of the 5810R Refrigerated Centrifuge
- Fast Cool function pre-cools chamber in as little as 15 minutes
- Temperature range from 9 °C to 40 °C
- Standby refrigeration keeps rotor chamber at set temperature while lid is closed
 POCD Scientific ~ Eppendorf’s Australian distributor for Multipurpose Centrifuges 5810 & 5810R ~ technical |
| Model |
Centrifuge 5810 |
| Dimensions (W x D x H in cm): |
54 x 61 x 35 |
| Weight: |
68kg |
| Max. power requirement: |
850 W |
| Power supply: |
230 V, 50-60 Hz |
 |
| Model |
Centrifuge 5810 R (refrigerated) |
| Dimensions (W x D x H in cm): |
70 x 61 x 35 |
| Weight: |
99kg |
| Max. power requirement: |
1.350 W |
| Power supply: |
230 V, 50-60 Hz |
Multipurpose Centrifuges 5810 & 5810R Braking times
| Rotor |
0 |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
8 |
9 |
| A-4-81 |
532 |
189 |
174 |
143 |
131 |
109 |
95 |
85 |
59 |
31 |
| A-4-81-MTP/Flex |
643 |
191 |
174 |
142 |
131 |
110 |
94 |
83 |
58 |
30 |
| A-4-62 |
740 |
190 |
170 |
140 |
130 |
110 |
95 |
85 |
55 |
26 |
| A-4-62-MTP |
620 |
190 |
170 |
140 |
130 |
110 |
95 |
85 |
55 |
26 |
| A-4-44 |
470 |
300 |
270 |
220 |
200 |
140 |
100 |
75 |
45 |
23 |
| A-2-DWP |
304 |
174 |
130 |
118 |
100 |
75 |
51 |
44 |
32 |
14 |
| T-60-11 |
800 |
280 |
140 |
95 |
70 |
55 |
45 |
40 |
36 |
36 |
| F-34-6-38 |
880 |
370 |
280 |
190 |
170 |
150 |
125 |
95 |
75 |
54 |
| F-45-30-11 |
240 |
140 |
70 |
45 |
35 |
30 |
25 |
22 |
19 |
18 |
| FA-45-30-11 |
240 |
140 |
70 |
45 |
35 |
30 |
25 |
22 |
19 |
18 |
| FA-45-48-PCR |
169 |
119 |
60 |
41 |
31 |
26 |
22 |
19 |
17 |
16 |
| The given times are to be considered guidelines. While level 9 means "strongest braking", level 0 means "free deceleration", whereby considerable fluctuations can occur with this uncontrolled deceleration depending upon the condition of the device and the load. The deceleration times for the 230 and 120 V devices are almost identical. |
 POCD Scientific ~ Eppendorf’s Australian distributor for Multipurpose Centrifuges 5810 & 5810R ~ rotors |
Rotors and Adaptors for Multipurpose Centrifuges 5804, 5804 R, 5810, and 5810 R

Eppendorf Multipurpose Centrifuges accommodate fixed-angle and swing-bucket rotors for a variety of applications. A large selection of Adaptors allows maximum application flexibility with a minimum number of rotors.
Product features
- All rotors, buckets, and Adaptors are autoclavable at 121°C
- Stackable modular Adaptors for swing-bucket rotors are easy to disassemble for vessels of different lengths, for cleaning and autoclaving
- Adaptors are color coded for quick and simple identification
- Buckets can be easily sealed with aerosol-tight covers to provide protection against biohazardous samples
|
Quick Find links:
|
|
|
for 5810 and 5810R Only
|
Swing-bucket rotor A-4-81 for Centrifuges 5810/5810 R
- High-capacity swing-bucket rotor with speed up to 3.220 x g (4.000 rpm)
- Ideal for medical and biotechnology laboratories: Broad range of adapters enable high versility for tubes and flasks from 0.2 mL to 500 mL
- Rectangular design of the buckets allows for higher tube capacity compared to standard round buckets
- Rotor A-4-81 can handle up to 28 x 50 mL Falcon® or 48 x 15 mL Falcon® tubes
- Rotor A-4-81 provides high capacity for up to 72 x 5/7 mL or 64 x 10 mL blood collection tubes
- Buckets, lids and adapters are autoclavable (121 °C, 20 min)
- Lids for aerosol-tight* centrifugation optionally available
®: Falcon is a registered trademark of Becton Dickinson.
* Aerosol-tightness is tested and certified by the Centre of Emergency Preparedness and Response, Health Protection Agency, Porton Down, Great Britain
Rotor model and accessories | Max. rotational speed (rpm) | Max. rcf (x g) | Order no. |
| A-4-81 |
| 4,000 | 3,250 | 5810 718.007 |
| 4,000 | 3,250 | 5810 743.001 |
| Rectangular buckets 400 ml |
| | 5810 730.007 |
| | 5810 724.007 |
| New accessoires 2011 |
| | 5825 730.003 |
| | 5820 718.005 |
| | 5825 707.003 |
|
| Adapter for 400 ml rectangular bucket for standard, Vacutainer®, Falcon®tubes & flasks |
| Tube size | Number of tubes | | Diameter of | Order no. |
| per adapter | per rotor | adapter bore (mm) | (set of 2 adapters) |
| 2.6-7 ml | 25 | 100 | 13 | 5810 720.001 |
| 7-17 ml | 16 | 64 | 17.5 | 5810 721.008 |
| 15 ml Falcon® | 12 | 48 | 17.5 | 5810 722.004 |
| 50 ml Falcon® | 5 | 20 | 31 | 5810 723.000 |
| 50 ml Centriprep® *2 | 5 | 20 | 31 | 5810 739.004 |
| 180-250 ml*2 | 1 | 4 | 62 | 5825 722.000 |
| 400 ml | 1 | 4 | 81 | 5810 728.002 |
| Wide-mouth flask, 400 ml, plastic |
| Set of 2 pieces | | | | 5810 729.009 |
|
*1 For the use of the rotor A-4-81 in older models of the centrifuge 5810/5810R an upgrade of the electronics my be necessary.
*2 Not for use with aerosol-tight caps.
®: Registered trademark of Becton Dickinson
|
| Adapters for swing-bucket rotor A-4-81 |
| |
| Tube |
 |
 |
 |
 |
 |
| Tube capacity (ml) |
2.6-7 |
7-17 |
15ml Falcon |
50ml Falcon |
400ml Flask** |
| * Max. tube length for all positions below the cap (mm) |
107 |
118 |
119 |
116 |
- |
| * Max. tube length for all positions without cap (mm) |
108 |
118 |
121 |
122 |
133 |
| Adapter |
 |
 |
 |
 |
 |
| Order number |
5810 720.001 |
5810 721.008 |
5810 722.004 |
5810 723.000 |
5810 728.002 |
| Number of tubes per adapter/per rotor |
25/100 |
16/64 |
12/48 |
5/20 |
1/4 |
| Adapter bore (mm) |
13 |
17.5 |
17.5 |
31 |
81 |
| Max. rcf (g-force) |
3,000 |
3,000 |
3,100 |
3,100 |
3,250 |
| Centrifugation radius (mm) |
168 |
168 |
173 |
173 |
180 |
|
Max. rotational speed: 4,000 rpm
Max. rotational speed for 4ºC (5810 R): 4,000 rpm
Temperature at max. rotational speed (5810 R): <8ºC (+23ºC room temperature and 1 hour run time)
* Exceeding the maximum tube length may reduce the loading capacity of the adapter. Please carry out a manual swing test.
** Available as a set of 2 flasks (see ordering information)
Adapters, caps and rotor are autoclavable: 20 min., 121ºC.
1) The total weight of a loaded rectangular bucket (incl. adapter and tubes) may not exceed 1.4 kg. The weight specified on the rotor applies to the rectangular bucket including load.
2) Swing-bucket rotors are delivered with rectangular buckets of one weight category. When placing follow-up orders for rectangular buckets, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In a rectangular bucket, the overall weight of adapters and test tubes must not exceed a weight of 800g.
®: Registered trademark of Becton Dickinson.
|
|
|
|
|
Swing-bucket rotor A-4-81-MTP/Flex for Centrifuges 5810 / 5810 R
Rotor A-4-81 can be upgraded with MTP/Flex buckets to spin plates
Max. rcf: 2.900 x g (4.000 rpm)
With a maximum loading height of 60 mm rotor A-4-81-MTP/Flex can handle up to
- 4 x 4 MTP
- 4 x 2 Cell culture plates
- 4 x 1 DWP
The open design of the MTP/Flex bucket without cross bar allows for easy loading and unloading
Rotor model and accessories | Max. rotational speed (rpm) | Max. rcf (x g) | Order no. |
| A-4-81-MTP/Flex |
| Swing-bucket rotor incl.4 microtest-plate buckets | 4,000 | 2,900 | 5810 725.003 |
| MTP/Flex bucket (set of 4) | 5810 741.009 |
| MTP/Flex bucket (set of 2) | 5810 742.005 |
|
Flex buckets (special MTP buckets) for A-4-81
- These special MTP buckets have a wider application range than the standard MTP buckets of the A-4-81-MTP
- Available as set of 2 or 4
- The adapters shown are exclusively designed for use with IsoRacks and cell culture flasks, as well as for the Perfectprep Plasmid 384 Kit
IsoRack adapter for MTP / Flex buckets
- A permanently safe centrifugation support is now available for the proven IsoRacks
- A total of up to 96 test tubes can be used in the swing-bucket rotor
- Two versions for 0.5 ml or for 1.5/2 ml test tubes available
- Max. rcf at 4,000 1/min: 2,600 x g
| Description | Order no. |
| Adapter for Flex buckets for the use of IsoRacks and cell culture flasks |
| Tube type/size | Number of tubes per adapter/rotor | set of 2 adapters |
| 0.5 ml tube in the IsoRack | 24 | 96 | 5825 708.008 |
| 1.5/2 ml tube in the IsoRack | 24 | 96 | 5825 709.004 |
| IsoRack starter set for Flex buckets* |
| 2 x IsoRack adapters, 2 x IsoRack with lid and 2 x cool packs (0°C IsoPack) | |
| for 0.5 ml and 1.5/2 ml micro test tubes | 5825 721.004 |
|
Slides adapter CombiSlide® for MTP buckets
- Safe centrifugation support for glass slides and microarrays in all MTP/DWP rotors in the catalogue
- Maximum of 12 slides per bucket
- Most extensive flexibility attained with help of novel slot frames, which allow centrifugation either across or parallel to the rotational direction
- Max. rcf: 1000 x g (corresponds to 770 1/min for MTP bucket of A-4-81 / A-4-62)
| Description | Order no. |
| Adapter for all MTP/DWP rotor buckets for the 5810 / 5810 R |
| Glass slides* | 12 | 48 | 5825 706.005 |
|
PCR plate adapter for MTP bucket
- Safe centrifugation of PCR plates (skirted, unskirted and semi-skirted) in all MTP/DWP rotors in the catalog
- Two versions for 96 and 384-well PCR plates available
- Optimized bores for most commonly available PCR plates
- Max. rcf in the A-4-81-MTP: 2,700 x g (at 4,000 1/min)
| Description | Order no. |
| Adapter for all MTP/DWP rotor buckets for the 5810 / 5810 R |
| 96-well PCR plate* | 1 | 4 | 5825 711.009 |
| 384-well PCR plate* | 1 | 4 | 5825 713.001 |
|
High-Speed Fixed-angle 30-place Microcentrifuge Rotors
- Spin up to 20,800 x g
- Quick release lid

Aerosol-tight Fixed-angle Rotor, for microcentrifuge tubes
Type FA-45-30-11

Fixed-angle Rotor,
for microcentrifuge tubes Type F-45-30-11 |
Specifications |
Catalog No. - Standard - Aerosol-tight |
022637002 022637100 |
| Maximum Capacity |
30 x 2.0 ml |
| Maximum Speed |
14,000 rpm |
| Maximum RCF |
20,800 x g |
| Maximum Radius |
95 mm |
| Maximum Vessel Diameter |
11.5 mm |
| Maximum Vessel Length |
50 mm |
Tube (Please observe the limiting data from manufacturer regarding the tubes maximum load- bearing capacity.) |
| |
 |
 |
 |
 |
| Tube capacity (ml) |
1.5-2 |
0.2 |
0.4 |
0.5 |
| Adaptor for 30 x 1.5 ml/2.0 ml fixed-angle rotor for use with PCR and micro test tubes |
| Adaptor |
Can be inserted without Adaptor |
Set of 6
 |
Set of 6
 |
Set of 6
 |
| Catalog No. |
- |
022636260 |
022636243 |
022636227 |
| Number of tubes per Adaptor/per rotor |
«/30 |
1/30 |
1/30 |
1/30 |
| Adaptor bore (dia. mm) |
- |
6 |
6 |
8 |
| Max. RCF (g-force) |
20,800 |
16,200 |
20,800 |
18,400 |
| Centrifugation radius (mm) |
95 |
74 |
95 |
84 |
- Max. rotational speed: 14,000 rpm
- Max. rotational speed for 4°C (5804 R/5810 R): 14,000 rpm
- Temperature at max. rotational speed (5804 R, 230 V/5810 R, 230 V): <4°C*
| * |
23°C room temperature and 1 hour run time.
|
 |
Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C. |
High-speed Fixed-angle Rotor for 6 x 85 ml Tubes
- Adaptors accommodate 1.5 to 50 ml tubes
- Spins conical tubes at high speeds

High-speed Fixed-angle Rotor, 6 x 85 ml
Type F-34-6-38 |
Specifications |
| Catalog No. |
022637207 |
| Maximum Capacity |
6 x 85 ml |
| Maximum Speed |
11,000 rpm* |
| Maximum RCF |
15,500 x g* |
| Maximum Radius |
115 mm |
| Maximum Vessel Diameter |
38 mm |
| Maximum Vessel Length |
120 mm |
| *12,000 rpm/18,500 x g in Model 5810 R. To maintain 4°C with this rotor, a maximum speed of 10,500 rpm is required. |
| Tube |
|
(Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.) |
 |
 |
 |
 |
| Tube capacity (ml) |
1.5-2 |
7-15 |
15 ml conical |
15-18 |
| Max. tube length for all positions below the rotor lid (mm)* |
110 |
112 |
123 |
123 |
| Adaptor for 6 ml x 85 ml fixed-angle rotor for use with Standard, conical and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
|
| Catalog No. |
022637215 |
022637223 |
022637274 |
022637231 |
| Number
of tubes per Adaptor/per rotor |
4/24 |
2/12 |
1/6 |
1/6 |
| Adaptor bore (dia. mm) |
11 |
16 |
17 |
18 |
| Max.
RCF (g-force) at 11,000 rpm |
15,290 |
15,150 |
14,470 |
14,750 |
| Max.
RCF (g-force) at 12,000 rpm - 5810 R |
18,190 |
18,030 |
17,230 |
17,550 |
| Centrifugation radius (mm) |
113 |
112 |
107 |
109 |
| Tube |
| (Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.) |
 |
 |
 |
 |
| Tube capacity (ml) |
20-30 |
50 |
50 ml conical |
85 |
| Max. tube length for all positions below the rotor lid (mm)* |
123 |
123 |
121 |
121 |
| Adaptor for 6 ml x 85 ml fixed-angle rotor for use with Standard, conical and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Can be inserted without Adaptor |
| Catalog No. |
022637240 |
022637258 |
022637266 |
- |
| Number of tubes per Adaptor/per rotor |
1/6 |
1/6 |
1/6 |
-/6 |
| Adaptor bore (dia. mm) |
26 |
29 |
29.5 |
38 |
| Max. RCF (g-force) at 11,000 rpm |
14,880 |
15,150 |
14,610 |
15,560 |
| Max. RCF (g-force) at 12,000 rpm - 5810 R |
17,710 |
18,030 |
17,390 |
18,500 |
| Centrifugation radius (mm) |
110 |
112 |
108 |
115 |
- Temperature at max. rotational speed (12,000 rpm), 5810 R (230 V): <10°C**
- Temperature at max. rotational speed (12,000 rpm), 5810 R (120 V): <12°C**
- Max. rotational speed for 4°C, 5810 R: 10,500 rpm
- Temperature at max. rotational speed (11,000 rpm), 5804 R: <6°C**
- Max. rotational speed for 4°C, 5804 R: 10,500 rpm
| * |
Exceeding the maximum tube length may reduce the loading capacity of the Adaptor. |
| ** |
23°C room temperature and 1 hour run time. |
 |
 |
| |
Adaptors, aerosol-tight covers, and rotors are autoclavable 20 min. at 121°C. |
Horizontal Drum Rotor
- Accepts up to 60 x 1.5 ml or 120 x 0.4 ml tubes for high throughput
- Carriers can be used as tube racks for easy handling

Drum Rotor,
Type T-60-11 |
Specifications |
| Catalog No. |
022638505 |
| Maximum Capacity |
60 x 1.5-2.0 ml, 120 x 0.4 ml |
| Maximum Speed |
14,000 rpm |
| Maximum RCF |
16,435 x g |
| Maximum Radius |
750 mm |
| Maximum Vessel Diameter |
11 mm |
| Maximum Vessel Length |
41 mm |
PCR/MTP/DWP Swing-bucket Rotor with Two Buckets
- Ideal for deepwell blocks and purification kit assemblies with heights greater than 45mm
- Use with Eppendorf Perfectprep® Plasmid 96 Spin Kits
- Spins up to 10 microplates or 4 deepwell plates

Deepwell plate Rotor, includes
2 Deepwell plate buckets
Type A-2-DWP |
Specifications |
| Catalog No. |
022638564 |
| Maximum Capacity |
2 Deepwell Assemblies (filter and catch-plate stacks) |
| Maximum Speed |
3,700 rpm |
| Maximum RCF |
2,250 x g |
| Maximum Radius |
147 mm |
| Maximum Vessel Dimensions, W x D |
91 x 134 mm |
| Maximum Height |
85 mm |
| Type |
Microplate |
Cell culture |
DWP |
Kits |
 |
 |
 |
 |
 |
| Number per bucket/per rotor |
5/10 |
4/8 |
2/4 |
1/4 |
| Max. RCF (g-force) |
2,250 |
2,250 |
2,250 |
2,250 |
| Max. Centrifugation radius (mm) |
147 |
147 |
147 |
147 |
| Dimensions: Length x Width x Height (mm) |
128 x 85.6 x 15 |
128 x 85.6 x 21 |
128 x 85.6 x 45 |
128 x 85.6 x 60 |
- Max. rotational speed: 3,700 rpm
- Max. rotational speed for 4°C (5804 R / 5810 R): 3,700 rpm
- Temperature at max. rotational speed (5804 R / 5810 R): <0°C**
| * |
The deepwell plate rotor 22 63 856-4 is delivered with buckets of one weight category. When placing follow-up orders for a deepwell plate bucket, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In the case of plates used in deepwell plate buckets for rotor 22 63 856-4 (commercially available microplates, culture plates, and deepwell plates), the maximum weight must not exceed 380 g and the maximum height must not exceed 85 mm. If plates other than the above-mentioned plates are used, the user must consult Eppendorf to confirm the suitability of using these plates. |
| ** |
23°C room temperature and 1 hour run time. |
| |
|
| |
Buckets and rotors are autoclavable 20 min. at 121°C. |
|
Swing-bucket Rotor with Four 100 ml Buckets
- Adaptors accommodate 0.25 to 100 ml tubes
- Accepts four 2 x 50 ml buckets to spin eight 50 ml conicals>
- Spins up to 4,500 x g for quick separation

Swing-bucket Rotor,for
standard tubes, includes 4 x 100 ml rectangular buckets
Type A-4-44 |
Specifications |
| Catalog No. |
022637401 |
| Maximum Capacity |
4 x 100 ml |
| Maximum Speed |
5,000 rpm |
| Maximum RCF |
4,500 x g |
| Maximum Radius |
161 mm |
| Maximum Vessel Diameter |
46 mm |
| Maximum Vessel Length |
125 mm |
| Tube |
|
|
 |
 |
 |
 |
 |
| Tube capacity (ml) |
3-5 |
1.5-2 |
4-7 |
3-15 |
7-17 |
| Typical tube size** |
10 x 75 |
11 x 38 |
12 x 75 or 13 x 100 |
16 x 100 |
17 x 100 |
| Max. tube length with aerosol-tight cover*** |
102 |
105 |
106 |
106 |
106 |
| Max. tube length without aerosol-tight cover*** |
105 |
111 |
108 |
108 |
110 |
| Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
|
Set of 2
 |
Set of 2
 |
Set of 2
|
Set of 2
 |
| Catalog No. |
022637509 |
022637525 |
022637541 |
022637568 |
022637584 |
| Number of tubes per Adaptor/per rotor |
14/56 |
12/48 |
9/36 |
7/28 |
6/24 |
| Adaptor bore (dia. mm) |
11 |
11 |
13 |
16 |
17.5 |
| Max. RCF (g-force) |
4,192 |
4,136 |
4,192 |
4,192 |
4,192 |
| Centrifugation radius (mm) |
150 |
148 |
150 |
150 |
150 |
| Tube |
 |
 |
For use with aerosol-tight cover
 |
|
 |
 |
| Tube capacity (ml) |
15 ml conical |
15 ml conical |
7-18 |
12-30 |
50 ml conical |
| Typical tube size** |
17 x 120 |
17 x 120 |
18 x 100 |
25 x 92 |
30 x 115 |
| Max. tube length with aerosol-tight cover*** |
109 |
121 |
104 |
100 |
112 |
| Max. tube length without aerosol-tight cover*** |
121 |
121 |
107 |
108 |
122 |
| Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
|
Set of 2
|
Set of 2
 |
Set of 2
|
Set of 2
|
| Catalog No. |
022637606 |
022637614 |
022637622 |
022637649 |
022637665 |
| Number of tubes per Adaptor/per rotor |
4/16 |
2/8 |
4/16 |
2/8 |
1/4 |
| Adaptor bore (dia. mm) |
17.5 |
17.5 |
20 |
26 |
31 |
| Max. RCF (g-force) |
4,332 |
4,388 |
4,192 |
4,192 |
4,332 |
| Centrifugation radius (mm) |
155 |
157 |
150 |
150 |
155 |
| Tube |
 |
For use with aerosol-tight cover
|
 |
 |
 |
|
| Tube capacity (ml) |
50 ml conical |
50 ml conical |
30-50 |
50-75 |
80-120 |
| Typical tube size** |
30 x 115 |
30 x 115 |
30 x 107 |
34 x 100 |
44 x 100 |
| Max. tube length with aerosol-tight cover*** |
119 |
- |
108 |
108 |
100 |
| Max. tube length without aerosol-tight cover*** |
119 |
119 |
122 |
119 |
114 |
| Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
 |
Bucket
|
Set of 2
|
Set of 2
|
Set of 2
 |
| Catalog No. |
022637673 |
022637452 |
022637681 |
022637703 |
022637720 |
| Number of tubes per Adaptor/per rotor |
1/4 |
2/8 |
1/4 |
1/4 |
1/4 |
| Adaptor bore (dia. mm) |
31 |
- |
31 |
35 |
45 |
| Max. RCF (g-force) |
4,388 |
4,500 |
4,192 |
4,192 |
4,192 |
| Centrifugation radius (mm) |
157 |
161 |
150 |
150 |
150 |
- Max. rotational speed: 5,000 rpm
- Max. rotational speed for 4°C (5804 R / 5810 R): 4,800 rpm
- Temperature at max. rotational speed (5804 R / 5810 R): 6°C****
| * |
Swing-bucket rotors are delivered with rectangular buckets of one weight category. When placing follow-up orders for rectangular buckets, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In a rectangular bucket, the overall weight of the Adaptors and test tubes used must not exceed 310 g. |
| ** |
Diameter x length in mm. |
| *** |
Exceeding the maximum tube length may reduce the loading capacity of the Adaptor. Please carry out a manual swing test. |
| **** |
23°C room temperature and 1 hour run time.
|
| |
Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C. |
PCR Strip Rotor
- Max. rotational speed 10,000 rpm
- Up to 6 x 8-tube strips, 6 x 5-tube strips or 48 x 0.2 ml PCR single tubes
- Safe and easy handling without additional lid
 Type F-45-48-PCR |
Specifications |
| Catalog No. |
022638581 |
| Maximum Speed |
12,000 rpm |
| Maximum RCF |
15,350 x g |
High-capacity Swing-bucket Rotor for Centrifuges 5810 and 5810 R Only - with Four 250 ml Buckets
- Adaptors accept up to 36 x 15 ml conical tubes
- Many different Adaptor types spin a variety of tubes per run
- Switch to microplate buckets for spinning deepwell plates and blocks (see below)
|

Swing-bucket Rotor,for standard tubes, includes 4 x 250 ml rectangular
buckets
Type A-4-62
|
Specifications |
|
With Four 250 ml Buckets |
| Catalog No. |
022638009 |
| Maximum Capacity |
4 x 250 ml |
| Maximum Speed (rpm) |
4,000 |
| Maximum RCF |
3,200 x g |
| Maximum Radius (mm) |
180 |
| Maximum Vessel Diameter (mm) |
62 |
| Maximum Vessel Length (mm) |
140 |
| Tube |
| |
 |
 |
 |
 |
 |
| Tube capacity (ml) |
3-5 |
1.5-2 |
4-7 |
3-15 |
7-17 |
| Typical tube size** |
10 x 75 |
11 x 38 |
12 x 75 or 13 x 100 |
16 x 100 |
17 x 100 |
| Max. tube length with aerosol-tight cover*** |
115 |
118 |
118 |
116 |
114 |
| Max. tube length without aerosol-tight cover*** |
123 |
122 |
121 |
121 |
118 |
| Adaptor for 250 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
| Catalog No. |
022638203 |
022638220 |
022638246 |
022638262 |
022638301 |
| Number of tubes per Adaptor/per rotor |
25/100 |
16/64 |
15/60 |
12/48 |
12/48 |
| Adaptor bore (dia. mm) |
11 |
11 |
13 |
16 |
17.5 |
| Max. RCF (g-force) |
3,072 |
3,036 |
3,072 |
3,072 |
3,072 |
| Centrifugation radius (mm) |
173 |
171 |
173 |
173 |
173 |
| Tube |
| |
 |
|
 |
 |
**
|
| Tube capacity (ml) |
15 ml conical |
7-18 |
12-30 |
50 ml conical |
50 ml conical |
| Typical tube size** |
17 x 120 |
18 x 100 |
25 x 92 |
30 x 115 |
30 x 115 |
| Max. tube length with aerosol-tight cover*** |
121 |
119 |
116 |
116 |
- |
| Max. tube length without aerosol-tight cover*** |
127 |
129 |
119 |
122 |
122 |
| Adaptor for 250 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
| Catalog No. |
022638289 |
022638327 |
022638360 |
022638343 |
022638351 |
| Number of tubes per Adaptor/per rotor |
9/36 |
8/32 |
4/16 |
3/12 |
4/16 |
| Adaptor bore (dia. mm) |
17.5 |
20 |
26 |
31 |
31 |
| Max. RCF (g-force) |
3,161 |
3,072 |
3,072 |
3,161 |
3,072 |
| Centrifugation radius (mm) |
178 |
173 |
173 |
178 |
173 |
| Tube |
| |
 |
 |
|
 |
|
| Tube capacity (ml) |
30-50 |
50-75 |
80-120 |
180-250 |
|
| Typical tube size** |
30 x 107 |
34 x 100 |
44 x 100 |
62 x 125 |
|
| Max. tube length with aerosol-tight cover*** |
113 |
118 |
125 |
127 |
|
| Max. tube length without aerosol-tight cover*** |
115 |
122 |
138 |
136 |
|
| Adaptor for 250 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
|
Set of 2
 |
Set of 2
|
|
| Catalog No. |
022638386 |
022638408 |
022638424 |
022638441 |
|
| Number of tubes per Adaptor / per rotor |
4 / 16 |
2 / 8 |
1 / 4 |
1 / 4 |
|
| Adaptor bore (dia. mm) |
31 |
35 |
45 |
62 |
|
| Max. RCF (g-force) |
3,072 |
3,072 |
3,072 |
3,200 |
|
| Centrifugation radius (mm) |
173 |
173 |
173 |
180 |
|
Max. rotational speed: 4,000 rpm Max. rotational speed for 4°C (5810 R): 4,000 rpm
Temperature at max. rotational speed (5810 R): <4°C****
| * |
Swing-bucket rotors are delivered with rectangular buckets of one weight category. When placing follow-up orders for rectangular buckets, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In a rectangular bucket, the overall weight of the Adaptors and test tubes used must not exceed 630 g. |
| ** |
Caution: When 50 ml Falcon® tubes are used in Adaptor 22 63 835-1, the aerosol-tight cap can no longer be attached. |
| *** |
Diameter x length in mm. |
| **** |
Exceeding the maximum tube length may reduce the loading capacity of the Adaptor. Please carry out a manual swing test. |
| ***** |
23°C room temperature and 1 hour run time. |
| |
Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C. |
With Four Microplate Buckets
- Spins up to 16 microplates per run for high throughput

Swing-bucket Rotor, includes
4 microplate buckets
Type A-4-62-MTP
|
Specifications |
| |
With Four Microplate Buckets |
| Catalog No. |
022638041 |
| Maximum Capacity |
16 Microplate or 4 Deepwell Blocks |
| Maximum Speed (rpm) |
4,000 |
| Maximum RCF |
2,750 x g |
| Maximum Radius (mm) |
154 |
| Maximum Height (mm) |
45 |
- Max. rotational speed: 4,000 rpm
- Max. rotational speed for 4°C (5810 R): 4,000 rpm
- Temperature at max. rotational speed (5810 R): <0°C**
| * |
Swing-bucket rotors are delivered with microplate buckets of one weight category. Whenplacing follow-up orders for a replacementmicroplate bucket (Catalog No. 22 63 807-6),please state the existing weight category, whichis printed on the inside of the groove (e.g. 86,with the last two figures in grams). In the caseof plates used in the microplate buckets ofSwing-bucket Rotor 22 63 800-9 and/or 22 63804-1 (commercially available microplates,culture plates, and deep-well plates), themaximum weight must not exceed 380 g and themaximum height must not exceed 45 mm.
If plates other than the above-mentionedplates are used, the user must consult Brinkmannto confirm the suitability of using theseplates.The Swing-bucket Rotor 22 63 800-9 can beloaded with rectangular buckets or microplatebuckets. If the Swing-bucket Rotor is alreadyavailable with rectangular buckets, fourmicroplate buckets may be ordered and insertedinto the same rotor cross, for user-friendlyexchange of rectangular buckets and microplatebuckets
|
| ** |
23°C room temperature and 1 hour run time. |
| |
Buckets and rotors are autoclavable 20 min. at 121°C. |
Super High-capacity Swing-bucket Rotor for Centrifuges 5810 and 5810 R
With four 400 ml buckets
- High capacity for standard tubes; modular Adaptors for bottles and tubes
- Accommodates up to 4 x 400 ml bottles, 48 x 15 ml conical tubes, or 100 x 7 ml tubesl Ideal for medical and biotechnology laboratories

Swing-bucket Rotor, for standard tubes, includes 4 x 400 ml rectangular buckets
Type A-4-81 |
Specifications |
| Catalog No. |
022638602 |
| Maximum Capacity |
4 x 400 ml |
| Maximum Speed |
4,000 rpm |
| Maximum RCF |
3,250 x g |
| Maximum Radius |
180 mm |
| Maximum Vessel Diameter |
81 mm |
| Maximum Vessel Length |
133 mm |
| Tube |
|
|
 |
 |
 |
 |
 |
| Tube capacity (ml) |
2.6-7 |
7-17 |
15 ml conical |
50 ml conical |
50 ml Centriprep®
or skirted conical
(not shown) |
| Max. tube length with aerosol-tight cover** |
107 |
118 |
119 |
116** |
- |
| Max. tube length without aerosol-tight cover** |
108 |
118 |
121 |
122 |
121 |
| Adaptor for Flex-buckets for use with IsoRacks and cell culture flasks |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
| Catalog No. |
022638700 |
022638726 |
022638742 |
022638769 |
022638904 |
| Number
of tubes per Adaptor/per rotor |
25/100 |
16/64 |
12/48 |
5/20 |
5/20 |
| Adaptor bore (dia. mm) |
13 |
17.5 |
17.5 |
31 |
31 |
| Max.
RCF (g-force) |
3,000 |
3,000 |
3,100 |
3,100 |
3,100 |
| Centrifugation radius (mm) |
168 |
168 |
173 |
173 |
173 |
Max. rotational
speed (4?°C are maintained securely in the 5810
R at a room temperature of 23°C) |
4,000 |
4,000 |
4,000 |
4,000 |
4,000 |
| Bottle |
| |
 |
 |
|
|
|
| Tube capacity (ml) |
180-2502 |
400 ml bottle
|
|
|
|
| Max. tube length with aerosol-tight cover** |
- |
- |
|
|
|
| Max. tube length without aerosol-tight cover** |
133 |
133 |
|
|
|
| Adaptor for Flex-buckets for use with IsoRacks and cell culture flasks |
| |
Set of 2
 |
Set of 2
 |
|
|
|
| Catalog No. |
022638921 |
022638785 |
|
|
|
| Number of tubes per Adaptor/per rotor |
1/4 |
1/4 |
|
|
|
| Adaptor bore (dia. mm) |
62 |
81 |
|
|
|
| Max. RCF (g-force) |
3,100 |
3,250 |
|
|
|
| Centrifugation radius (mm) |
173 |
180 |
|
|
|
Max. rotational speed (4?°C are maintained securely in the 5810 R at a room temperature of 23°C) |
4,000 |
4,000 |
|
|
|
- Max. rotational speed: 4,000 rpm
- Max. rotational speed for 4°C (5810 R): 3,900 rpm
- Temperature at max. rotational speed (5810 R): <8°C***
| * |
Swing-bucket rotors are delivered with buckets of one weight category. When placing follow-up orders for buckets, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In a bucket, the overall weight of the Adaptors and test tubes used must not exceed 800 g. |
| ** |
Exceeding the maximum tube length may reduce the loading capacity of the Adaptor. Please carry out a manual swing test. |
| *** |
23°C room temperature and 1 hour run time. |
| |
|
| |
Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C. |
With Four Microplate Buckets
- Accommodates up to 16 microplates, 2 culture plates,
4 deepwell plates, or Perfectprep® Plasmid purification kit assemblies
- Swing-bucket Rotor with microplate buckets;max. rotational speed: 4,000 rpm (max. RCF: 2,900 x g)

Swing-bucket Rotor, includes
4 microplate flex buckets
Type A-4-81-MTP |
Specifications |
| Catalog No. |
022638807 |
| Maximum Capacity |
2 Deepwell Assemblies (filter and catch-plate stacks) |
| Maximum Speed |
4,000 rpm |
| Maximum RCF |
2,900 x g |
| Type |
Microplate |
Cell culture |
DWP |
Kits |
| |
 |
 |
 |
 |
| Number per bucket / per rotor |
4 / 16 |
2 / 8 |
1 / 4 |
1 / 4 |
| Max. RCF (g-force) |
2,900 |
2,900 |
2,900 |
2,900 |
| Max. Centrifugation radius (mm) |
163 |
163 |
163 |
163 |
Max. rotational speed
(4°C are maintained securely in the 5810 R at a room temperature of 23°C) |
4,000 |
4,000 |
4,000 |
4,000 |
| Dimensions: Length x Width x Height (mm) |
128 x 85.6 x 15 |
128 x 85.6 x 21 |
128 x 85.6 x 45 |
128 x 85.6 x 60 |
- Max. rotational speed: 4,000 rpm
- Max. rotational speed for 4°C (5810 R): 4,000 rpm
- Temperature at max. rotational speed (5810 R): <0°C (+23°C room temperature and 1 hour run time)
|
| Adaptor for Flex-buckets for use with IsoRacks and cell culture flasks |
| |
 |
 |
|
|
| Tubes per Adaptor/per rotor |
24 / 96 |
24 / 96 |
|
|
| Max. RCF (g-force) |
2,700 |
2,600 |
|
|
| Max. Centrifugation radius (mm) |
150 |
146 |
|
|
Max. rotational speed
(4°C are maintained securely in the 5810 R at a room temperature of 23°C) |
4,000 |
4,000 |
|
|
- To use Rotors A-4-81/A-4-81-MTP with Centrifuge 5810/5810 R that were built prior to 2002, a minor electrical modification may be necessary.
- In the A-4-81-MTP bucket, plates should not exceed a weight of 380 g and a total height of 53 mm or 60 mm, respectively
|
Deepwell plate Rotor A-2-DWP-AT
For use in Centrifuge 5810/5810 R only
- Rotor can handle up to
- 2 × 4 MTP
- 2 × 2 Cell culture plates
- 2 × 1 DWP
- Maximum loading height of 60 mm
- Aerosol-tight lids*
- Max. rcf: 3,486 × g (4,500 rpm)
- Rotor and buckets are autoclavable (121 °C, 20 min)
* Aerosol-tightness is tested and certified by the Centre of Emergency Preparedness and Response, Health Protection Agency, Porton Down, Great Britain
Please note that a software upgrade will be necessary for older centrifuges when using Rotor A-2-DWP-AT. Older centrifuges can be retrofitted as of serial number 10,000 (for Centrifuges 5810) and as of serial number 12,000 (for Centrifuges 5810 R). Please contact your local Eppendorf organization or dealer for further information. All centrifuges with serial numbers > 60,000 already include the new software.
 POCD Scientific ~ Eppendorf’s Australian distributor for Multipurpose Centrifuges 5810 & 5810R ~ accessories |

Eppendorf’s rolling cabinet gives you the freedom and flexibility to use your Centrifuge 5804/ 5804 R or 5810/ 5810 R wherever you need it.
Space is often a limited factor in any lab. Our rolling cabinet enables you to use your existing space more efficiently.
Set up your Centrifuge 5804/R or 5810/R on the rolling cabinet if there isn’t enough space on the regular lab bench or use the rolling cabinet to transport and use the centrifuge in other labs.
Key product features rolling cabinet for Eppendorf Multiporouse Centrifuges
- One size for all 58xx Centrifuge models
- Very sturdy design for safe centrifugation
- Made of powder coated steel for high chemical resistance
- Storage for rotors and accessories
- Dampened drawer slide
- Inlaid with slip-resistant material
- Rolling cabinet with centrifuge fi ts under most lab benches
| Dimensions (Width / Height / Depth) |
70.0 / 45.0 / 61.5 cm |
| Depth with open drawer |
101.5 cm |
Max. loading of rolling cabinet
(incl. content of drawer) |
140 kg |
| Max. loading of drawer |
20 kg |
| Material |
Powder coated steel |
| Mobility |
Four castors, of which two are lockable
|
| Storage |
One drawer, equipped with slip-resistant material |
| Order no. |
5821 020.010 |

Rolling cabinet with open drawer

Drawer includes inlaid with slip-resistant material for perfect storage for rotors and accessories

Adjustable metal feet allow perfect stand for each Eppendorf Centrifuge 58xx

Four castors, of which two are lockable
 POCD Scientific ~ Eppendorf’s Australian distributor for Multipurpose Centrifuges 5810 & 5810R ~ downloads |
| Document |
Size |
download |
| 5840/5804R and 5810/5810R Operating Guide |
556 KB |
 |
| 5810/5810R 20% Bonus Giveaway Offer - download the flyer |
275 KB |
 |
| Supplementary Safety_precautions for loading buckets |
641 KB |
 |
| CombiSlide Adapter Supplementary sheet |
144 KB |
 |
| Instructions for using 400ml Wide-Neck Bottles |
59 KB |
 |
| Supplementary safety notes |
641 KB |
 |
| Rolling Cabinet Manual |
838 KB |
 |
| Eppendorf rotors - Usage instructions |
838 KB |
 |
These are PDF files, to obtain a free Adobe PDF reader:  |
 POCD Scientific ~ Eppendorf’s Australian distributor for Multipurpose Centrifuges 5810 & 5810R ~ FAQs |
| How are the Centrifuges 5804/5804 R and 5810/5810 R protected during transport? Is there a transport protection device? |
 |
'No, the Centrifuges 5804/5804 R and 5810/5810 R do not come with a transport protection device. They are packaged in polystyrene for protection. Because the rotor is not in the centrifuge during transport, there are no movable parts to make a transport protection device necessary.
| Is it possible to centrifuge 50 ml and 15 ml tubes together in one rotor? |
 |
Yes, it is possible, provided that the rotor is loaded symmetrically. This means that the 50 ml tubes and the 15 ml tubes are positioned opposite one another and are positioned symmetrically. Please use the appropriate adapters for the tubes.
| Can I still change the settings when I have started the Centrifuges 5804/5804 R or 5810/5810 R? |
 |
Yes, the speed, the time and the temperature can be changed even when the centrifuge is running. To do this, please press the SPEED, TIME or TEMP key and change the values by pressing the arrow keys.
| Do the adapters for the 0.2/0.4/0.5/0.6 ml tubes, which are designed for the rotors F-45-30-11 and FA-45-30-11, also fit into the rotor F-34-6-38 with the adapter 5804 770.005 (for 1.5/2.0 ml tubes)? |
 |
Yes, these adapters can also be used in the rotor F-34-6-38 with the adapter 5804 770.005.
| Does Eppendorf offer a rotor for PCR strips for Centrifuges 5804/5804 R or 5810/5810 R? |
 |
The PCR strip rotor F-45-48-PCR is available under the order number 5804 735.005. It can centrifuge up to 6 x 8-well strips at 12,000 rpm / 15,350 x g.
| In which rotor is it possible to centrifuge 50 ml DIN glass tubes? |
 |
50 ml DIN glass tubes (diameter 34 mm, length 100 mm) can be centrifuged in the adapter for 50 - 75 ml tubes (order number 5804 760.000) of the swing-bucket rotor A-4-44. Depending on the length of the tube, the uppermost module of the adapter can be removed for centrifugation of 50 ml DIN glass tubes.
| Is it possible to adjust the Centrifuge 5804/5804 R or 5810/5810 R so that it stops on its own, meaning without active braking? |
 |
Yes. Free deceleration of your centrifuge can be achieved by setting the braking level to 0.
| The Centrifuges 5804/5804 R and 5810/5810 R are programmable. What is saved during the programming? |
 |
All data is saved that can also be entered manually, such as rotational speed or RCF, maximum radius, run time, acceleration and braking ramps, "at set rpm", temperature.
| "Clean rotor" appears on the display of my centrifuge. Do I now need to remove the rotor from the centrifuge, and, if yes, how? |
 |
This display appears every 200 centrifugation runs. For cleaning the rotor and the rotor chamber, the rotor must first be removed from the centrifuge. To do this, loosen the rotor nut by turning it counter-clockwise with the supplied rotor key. Please clean the rotor and rotor chamber only with mild cleaning agents (e.g. 70% alcohol solutions or commercially available alcoholic disinfectants).
| How can I recognize in the display of the Centrifuges 5804/5804 R or 5810/5810 R whether I have selected "rpm" or "rcf"? |
 |
For rcf, a "*" is displayed in front of the value. rpm is not indicated by a symbol.
| Millipore offers centrifugation filter tubes (Centripreps). Can these tubes be centrifuged in a rotor of the Centrifuge 5804/5804 R or 5810/5810 R? |
 |
Yes, for 50 ml Centripreps we offer the adapter 5810 739.004 for the rotor A-4-81.
| What is the minimum rotational speed (rpm) of the Centrifuge 5804/5804 R or 5810/5810 R, and in which increments can it be increased? |
 |
The minimum rotational speed is 200 rpm, irrespective of the inserted rotor. The minimum rotational speed can be increased in increments of 10 rpm up to 5,000 rpm and then in increments of 100 rpm.
| Into which rotor of the Centrifuge 5804/5804 R or 5810/5810 R do 15 ml and 30 ml Corex tubes fit? |
 |
With a rubber adapter (outer diameter 18 mm) available from the tube manufacturer, the 15 ml Corex tubes fit into the adapter 5804 772.008 of the rotor F-34-6-38. With a rubber adapter (outer diameter 29 mm) available from the tube manufacturer, the 30 ml Corex tubes fit into the adapter 5804 774.000 of the rotor F-34-6-38. This rotor can be inserted into the Centrifuge 5804/5804 R or 5810/5810 R.
| How big is the g-force difference between the lowest and the highest plate when loading a bucket of the rotor A-2-DWP with four microplates? |
 |
The following g-forces are reached with the rotor A-2-DWP:
Lowest plate: 2,250 x g
Highest plate: 1,825 x g
| Is there an adapter for slides? |
 |
Up to 12 slides can be inserted into the CombiSlide adapter and then centrifuged in the swing-bucket rotors of Centrifuge 5804/5804 R or 5810/5810 R.
| Do 4 microplate buckets (order number 5810 702.003) also fit into the rotor A-4-62 (5810 709.008)? |
 |
Yes. It is the same rotor cross. Please note that mixed loading of the rotor with rectangular buckets and MTP buckets is not permitted.
| Can the rotor A-4-81 be operated with rectangular buckets and MTP / Flex buckets? |
 |
Yes, this so-called mixed loading is possible. Please ensure that the same type of buckets are always used in opposing positions.
| There are 2 different adapters for the 15 ml tubes. One for tubes with round bottoms and one with a conical bottom for Falcon tubes. Can I not use one adapter for both tubes? |
 |
No, unfortunately this is not possible. In order to be able to reach the high rotational speeds, the adapters must be form-locking.
| Do the adapters for the 0.5 ml tubes fit into the rack for 1.5/2.0 ml tubes of the rotor T-60-11? |
 |
Yes, the adapters can be inserted into the rotor for centrifugation of 0.5 ml tubes.
| The writing on one rotor of the Centrifuge 5804/5804 R or 5810/5810 R says, for example, "6x125g". Does this mean that the centrifuge only enables an RCF of 125 or that it can only be loaded with a maximum of 6 times 125 gram? |
 |
This label states that the rotor can be loaded with a maximum of 6 x 125 gram. "g" is the official abbreviation for "gram". Where "g" is used as an abbreviation of "g-force", the number and "x g" are used. Example: 10,800 x g.
| How many times can aluminum rotors be autoclaved? |
 |
Aluminum rotors can be autoclaved up to 100 times for 20 minutes at 121 °C.
|